Skip Navigation

European Journal of Heart Failure 2004 6(5):627-634; doi:10.1016/j.ejheart.2003.12.021
This Article
Right arrow Full Text Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Kjellström, B.
Right arrow Articles by Ryden, L.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kjellström, B.
Right arrow Articles by Ryden, L.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 2004 European Society of Cardiology

Six years follow-up of an implanted SvO2 sensor in the right ventricle*

Barbro Kjellströma,*, Cecilia Lindeb, Tom Bennetta, Åke Ohlssonc and Lars Rydenb

a Heart Failure Management, Medtronic Inc. MS CW210, Minneapolis, MN 55432, USA
b Karolinska Hospital Stockholm, Sweden
c Southern Hospital Stockholm, Sweden

* Corresponding author. Tel.: +1-673-514-8219; Fax: +1-763-514-8347. E-mail address: barbro.kjellstrom{at}medtronic.com


   Abstract

Introduction.: Mixed venous oxygen saturation (SvO2) is a standard invasive measure used in the management of congestive heart failure patients. The reliability of a long-term SvO2 sensor remains unproven.

Methods.: Nine patients (NYHA Class I/II, n=2/7) were implanted with a dual chamber pacemaker modified to utilize a right ventricular SvO2 lead (Medtronic Inc., Models 8007/4327A IPG/Lead). Invasive studies compared sensor SvO2 to reference (Optical Swan–Ganz catheter) at 0, 3 and 9 months. Symptom limited tests (Bikemax) with metabolic assessment and arterial oxygen saturation measurements performed 1–7 days, 3.5 and 9.5 months post-implant allowed for cardiac output calculations. Long-term sensor performance was confirmed by submaximal tests, Bikesubm in years 1–3, and Walkin-place every 6 months for the duration of follow-up.

Results.: Sensor SvO2 readings were stable over time when compared to the Swan–Ganz Catheter. Non-invasive CO measured during Bikemax was in normal ranges for this patient population, 3.7±0.9 l/min at rest and 8.4±2.2 l/min at peak-exercise. Resting SvO2 values from Bikesubm and Walkin-place did not change significantly over time (P0.1 vs. 1 year) and neither did the change from rest to peak exercise during Bikesubm (P0.05 vs. 1 year) or Walkin-place (P0.05 vs. 4 year).

Conclusion.: While limited in size, this small pilot study suggests that long-term monitoring of SvO2 by implanted devices may be feasible. The clinical value remains to be proven in future studies.

Key Words: Abbreviations • CO, cardiac output • PA, pulmonary artery • RV, right ventricle • SaO2, arterial oxygen saturation • SvO2, mixed venous oxygen saturation • S.D., standard deviation • VO2, oxygen uptake

Received June 26, 2003; Revised October 16, 2003; Accepted December 10, 2003


{star} Supported by Medtronic, Bakken Research Center, Maastricht, The Netherlands and the Swedish Heart and Lung Foundation.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?




Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.